22b4.2 testing the balance, 22b 4.2, Testing the balance – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 57: 2 testing the balance, Reference

Advertising
background image

RL-TS Series Operation Manual 53

22B

4.2 Testing the Balance

The amount of offset with respect to the reference weight can be checked. This is called “span test.”
Performing a span test will not calibrate the balance.

Use a weight for calibration that weighs 50% of the weighing capacity or heavier. To calibrate more
accurately, use a weight that is equivalent to the weighing capacity.

Please contact us if you wish inquire about or place an order for calibration weights.

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After “Func” is

displayed, release the finger.

2

Set to the span test mode.

Press the [Function] key several times to select “7.CA.”
Press the [Zero/Tare] key several times to select “7.CA 4.”

3

Save the setting.

Press the [Target] key.
The balance goes back to normal display.

4

Start span test.

<Press and hold>

Keep holding down the [Function] key until “tESt” is

displayed.
After zero-point testing (“on 0” blinks), the display changes

to “on F.S.”

5

Place the weight on the pan.

After “on F.S.” is displayed, place the weight on the pan.
The test starts.

6

The error is displayed.

The display first indicates “dIFF” and then the weight error

of the balance.
A weight error is as follows:
Weight error = True value – Current weight
When a weight error is more than “0”, the displayed weight is

less than the actual weight.
Press any key to return to normal display.