RL-TS Series Operation Manual 83

Problem

Cause

Remedy

No outputs

Intended output function settings
are not established.

Make

the

balance

compatible

in

communication conditions with the external
device by referring to their operation
manuals.

Check outputs with the following methods:

Initialize the function settings by referring to
“Section 7-3: Initializing (P.

X

83

X

) or Function

Setting List.

Then, set to a communication condition of
1,200 bps, 8-bit data, 2-bit stop bit, and no
parity by referring to the operation manual of
the external device.

Now, output is done once after stabilization,
after the [P] key is pressed. Check output by
pressing the [P] key.

Communication

conditions

disagree with the external device.

Wrong cable connection

To connect the balance to a general-use PC,
a cross cable is required (Refer to “Section
6-2: Connecting to External Devices via
RS-232C Interface” (P.

X

67

X

)).

Arrange it yourself or contact our Sales
Office.

The cable is disconnected or not
properly connected.

Check for proper cable connection.

The current settings of the
balance are unknown.

You can initialize the balance (“Section 7-3:
Initializing” (P.

X

83

X

)).

34B