Reference – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 39

Advertising
background image

RL-TS Series Operation Manual 35

6

Save the water temperature.

Press the [Target] key.

7

Hang only the cage to zero the weight

readout.

Hang only the cage on the underfloor weigher fixture.
Press the [Zero/Tare] key.
The readout becomes zero.


8

Measure the weight of the sample.

Place the sample on the cage.
After the weight display is stable, press the [Target] key to save

the weight of the sample in the air.
When the weight is saved, “◄” is displayed in the lower left of

the display.
Measuring a sample by placing it on the pan is also allowed.

9

Set the liquid container.

Set the container with water under the balance and submerge

only the cage (do not place the sample).
Press the [Zero/Tare] key to zero the readout.
This is to remove any residual error due to the cage.


10

Place the sample on the cage.

Put the sample on the cage and submerge the entire volume.
After the weight display is stable, press the [Target] key.
The measured specific gravity value is displayed.

Do not allow the cage to touch the liquid container.

Press the [Target] key to return to showing the sample’s normal

weight.