52 RL-TS Series Operation Manual
6
Place the weight on the pan.
After zero-point adjustment finishes and the display changes to
“on F.S.,” place the weight on the pan.
Adjustment starts.
7
Span adjustment starts.
After display changes to the blinking of “on F.S.” span
adjustment starts.
8
Span adjustment finishes.
After span adjustment finishes, “End” is displayed.
The balance goes back to normal display.
- Pressing any other key than the [Function] key cancels adjustment.
- The “1 - Err” display indicates that you used a weight weighing less than 50% of the
weighing capacity.
- The “2 - Err” display indicates that an error over 1.0% was detected, or the balance failed.