52 RL-TS Series Operation Manual

6

Place the weight on the pan.

After zero-point adjustment finishes and the display changes to

“on F.S.,” place the weight on the pan.
Adjustment starts.

7

Span adjustment starts.

After display changes to the blinking of “on F.S.” span

adjustment starts.

8

Span adjustment finishes.

After span adjustment finishes, “End” is displayed.
The balance goes back to normal display.

- Pressing any other key than the [Function] key cancels adjustment.
- The “1 - Err” display indicates that you used a weight weighing less than 50% of the

weighing capacity.

- The “2 - Err” display indicates that an error over 1.0% was detected, or the balance failed.

For more information, refer to “Section 7-2: Troubleshooting” (P.