5 id no. setting – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 64

Advertising
background image

60 RL-TS Series Operation Manual

5.5 ID No. Setting

Results of successfully completed span adjustments and span tests can be printed by ISO/GLP/GMP
compliant devices. Set the ID numbers printed together at this printing.
In situations such as when the same model is used, you can assign numbers that allow you to control them
easily.
When an ID number is set, the “◄” and “” indicators located in the upper left of the display light up.

You can use up to six digits in an ID number. The characters that you can use are as follows:
Space (blank), 0 to 9, A to F, -

1

Set to the function 2 setting mode.

+

Press the [Function] key while pressing the [Zero/Tare] key.

Release the fingers when “Func 2” is displayed.
The function 2 setting mode is set, displaying

“1. 1d0.”

2

Set the ID number.

Select “1” by pressing the [Zero/Tare] key.

3

Set to the ID number setting mode.

Press the [Function] key.

4

Enter the ID number.

Enter the ID number with the following steps:

1. Press the [Zero/Tare] key.

The leftmost digit blinks.

2. Select a character by pressing the [Zero/Tare] key.

Pressing the key toggles between space, 0 to 9, A to F, and

– (minus).

3. Pressing the [Function] key blinks the next digit.

Set the ID number by repeating steps 2 and 3.

Advertising