82 RL-TS Series Operation Manual

Problem

Cause

Remedy

The weight used for calibration
is slightly different in mass from
the weight used for checking.

Use the same weight during calibration and
checking.

The display does not move with
the M sign flashing.

(When the [Zero/Tare] key is
pressed, during a sampling in
parts counting mode, etc.)

The balance may be affected by
an external influence such as
wind and vibration.

The balance may be affected by an external
influence such as wind and vibration. Take
remedial actions or relocate the balance by
referring to the section “Do not use the
balance in the following places” (P. v).

The icon blinks when the balance
is operated on batteries.

The batteries are exhausted.

Replace the batteries.

The display is turned off when
the balance is operated on
batteries.

The display was turned off by the
auto power off function.

The auto power off function is activated if
the balance is left unused with no
measurement taken for approximately five
minutes.

Disable the auto power off function if it
interferes with your use of the balance.