BINDER KBF LQC 240 User Manual

Page 100

Advertising
background image

KBF LQC (E5.3) 07/2014

page 100/114


Fault description

Possible cause

Required measures

Controller (continued)

Ramp temperature transitions are

only realized as steps.

When using the Program Editor

of the software APT-COM™ 3

DataControlSystem, the setting

“step” has been selected.

Select setting “ramp” in the Pro-

gram Editor of the software APT-

COM™ 3 DataControlSystem

and transfer a program to the

chamber controller.

Humidity alarm message when

operating without humidity (hu-

midity switch (4) OFF)

Humidity set-point set to a value

> 0% r.F.

Manual Mode: Enter a humidity

set-point 0% r.H.
Program Mode: Enter a humidity

subprogram with humidity set-

point 0% r.H.

RESET button does not cancel

the notifying or alarm indication.

Cause of disturbance not re-

moved correctly
The RESET button permits reset-

ting notifying or alarm messages

for temperature and humidity only

with in a tolerance sector of +/- 2

°C resp. +/- 5 % r.H.

Remove cause of disturbance. If

the RESET button still does not

cancel the indication, contact

BINDER service.

Display flashing:

1999 or -1999 or 9999.

Sensor rupture between sensor

and controller or Pt 100 sensor

defective.

Contact BINDER service.

Short-circuit.
Initialization problem due to turn-

ing on the chamber too early.

Observe a delay time of about

30s between turning Off and On

again the chamber.

Miscellaneous
Fluorescent tube of ICH compli-

ant illumination does not illumi-

nate.

Defective fluorescent tube.

Replace the fluorescent tubes.

Impaired valve function of hose

burst protection.

Calcification.

Remove calcifications by citric

acid or acetic acid solutions

(chap. 4.3.4).
Inspection of the valve by a

plumber.

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.