You can order transport packing for transportation purposes from BINDER service.
Permissible ambient temperature range during transport:

If the steam humidifying system has NOT been emptied: +3 °C / 37.4°F to +60 °C / 140°F.

After BINDER Service has emptied the steam humidifying system: -10 °C / 14°F to +60 °C / 140°F.

With temperatures below +3 °C / 37.4°F, water must be completely removed from the humidifying sys-

tem.

CAUTION

Transport below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.

Contact BINDER Service before any transportation below +3 °C / 37.4°F.

3.3 Storage

Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-

rary decommissioning (chap. 19.2).
Permissible ambient temperature range during storage:

If the steam humidifying system has NOT been emptied: +3 °C / 37.4°F to +60 °C / 140°F.

After BINDER Service has emptied the steam humidifying system: -10 °C / 14°F to +60 °C / 140°F

With temperatures below +3 °C / 37.4°F, water must be completely removed from the humidifying sys-

tem.

CAUTION

Storage below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.

Contact BINDER Service before any transportation below +3 °C / 37.4°F.

Permissible ambient humidity: max. 70 % r.H., non-condensing

CAUTION

Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H.

for a long period.

Dry the appliance completely before shut-down:

Set the humidity to 0 % r.H. and turn on the humidity switch (4).

Set the temperature set point to 60 °C / 140°F for approx. 2 hours (Manual mode).

Only then, shut down the unit at the main power switch (3) and close the tap of the

water supply.

When after storage in a cold location you transfer the unit to its warmer installation site, condensation

may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and

is completely dry.
In case of a prolonged temporal decommissioning: Leave the unit door open or remove the access port

plugs.