10 binder pure aqua service (option), Maintenance, cleaning, and service, 1 maintenance intervals, service – BINDER KBF LQC 240 User Manual

Page 89: Danger

Advertising
background image

KBF LQC (E5.3) 07/2014

page 89/114

17.10 BINDER Pure Aqua Service (option)

The optional BINDER water treatment system (disposable system) serves to treat tap water. The lifetime

depends on water quality and the amount of treated water. The measuring equipment to assess the water

quality is reusable.

For detailed information on operating the water treatment system BINDER Pure Aqua Ser-

vice and its function, please refer to the operating manual Art. No. 7001-0269, delivered

with BINDER Pure Aqua Service.


18. Maintenance, cleaning, and service

18.1 Maintenance intervals, service

DANGER

Electrical hazard.
Danger of death.

The unit must NOT become wet during operation or maintenance works.

Do NOT remove the rear panel of the unit.

Before conducting maintenance work, turn off the unit at the main power switch and

disconnect the power plug.

General maintenance work must be conducted by licensed electricians or experts

authorized by BINDER.

Maintenance work at the refrigeration system must only be conducted by qualified

personnel who underwent training in accordance with EN 13313:2010 (e.g. a refriger-

ation technician with certified expert knowledge acc. to regulation 303/2008/EC). Fol-

low the national statutory regulations.

Ensure regular maintenance work is performed at least once a year and that the legal requirements are

met regarding the qualifications of service personnel, scope of testing and documentation. All work on the

refrigeration system (repairs, inspections) must be documented.

The warranty becomes void if maintenance work is conducted by non-authorized personnel.

Have conducted regular maintenance work on the steam humidifier at least once a year. The operating

behavior and the maintenance intervals of the humidifier essentially depend on the available water quality

and the amount of steam produced in the meantime.

We recommend cleaning the condensers at least twice a year. A qualified technician must

perform cleaning.

Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.

With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing)

several times a year.

Advertising