8 performance after power failure in program mode – BINDER KBF LQC 240 User Manual

Page 56

Advertising
background image

KBF LQC (E5.3) 07/2014

page 56/114

10.7 Repetition of a section or several sections within a program

The example of a set-point ramp temperature program of chap. 10.5 will serve as an example. The shad-

ed sections 02 and 03 shall be repeated e.g. 30 times.

Program

section

Set-point

temp.

Fan Section time Operation

lines

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

40

100 % 00:30:00

Off

1

0

-1999

+9999

02

60

100 % 01:30:00

Off

1

0

-5

+5

03

90

100 % 01:00:00

Off

1

0

-2

+2

04

90

100 % 03:20:00

Off

1

0

-1999

+9999

05

20

100 % 00:00:01

Off

1

0

-1999

+9999

The following table shows the resulting program, whereby the differences to the table above are shaded.

Program

section

Set-point

temp.

Fan Section time Operation

lines

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

40

100 % 00:30:00

Off

1

0

-1999

+9999

02

60

100 % 01:30:00

Off

1

0

-5

+5

03

90

100 % 01:00:00

Off

2

30

-2

+2

04

90

100 % 03:20:00

Off

1

0

-1999

+9999

05

20

100 % 00:00:01

Off

1

0

-1999

+9999

Sections 02 and 03 will be executed in total 31 times; only then, the program will continue.
Entry of the values into the display program table:

To have sections repeated infinitely, enter the number of cycles “Cy” as -1.

10.8 Performance after power failure in Program Mode

Execution of the program is resumed at the point where the interruption occurred with the latest set-

points reached during program course. The power failure is noted in the event list. No error message is

displayed indicating that a power failure has taken place.