2 messages concerning the humidity system – BINDER KBF LQC 240 User Manual

Page 69

Advertising
background image

KBF LQC (E5.3) 07/2014

page 69/114


When at least one of the light sensors is plugged-in, the maximum temperature of the chamber is limited

to 60 °C. If this temperature is reached due to an excessively high temperature set-point or to an error,

the message “

LIGHT SENSOR 60 °C!” is displayed. As soon as the unit has cooled down to a value of

59 °C or below, the message disappears.

12.2 Messages concerning the humidity system

Notification of the humidity system

The flashing blue notification “HUMID SYSTEM” appears on the controller. This message flashes with

a defined number of flashes from which the cause of the fault can be derived: number of flashes (each

one 1 sec on / 1 sec off) followed by a 5 second pause.

Error of the humidity system

First the blue notification “HUMID SYSTEM” appears on the controller. After 1 minute it changes to the

alarm message “HUMID SYSTEM”, and the buzzer is activated. After one minute, the alarm message

and buzzer can be reset with the RESET key.
Following this reset, the flashing blue notification “HUMID SYSTEM” appears on the controller. This

message flashes with a defined number of flashes from which the cause of the fault can be derived:

number of flashes (each one 1 sec on / 1 sec off) followed by a 5 second pause.

Number of flashes

Notification and measures

1

Purging is required soon. Turn off and on again the humidity switch to start

purging. After a successful purging the notification resets automatically.

2

Maintenance of the humidity system is required. Contact BINDER service.

Number of flashes

Cause of the error and measures

3, 4, 13

In case of freshwater supply via water pipe:

The water tap is closed, or the unit is defective.
In case of freshwater supply via freshwater can (option, chap. 17.9):

Water can is empty. Humidification is turned off. In case of refrigerating opera-

tion, the interior is strongly dehumidified. When the water supply is functional

again, the humidity system restarts, or the unit is defective.

8, 9

Check the length and location of the wastewater tube. If appropriate contact

BINDER service.

Other

The unit is defective. Contact BINDER service.


If the blue notification “HUMID SYSTEM” would be overlaid by the notification or alarm message “HUMID

RANGE”, you can set the humidity set-point to the actual value and then reset the “HUMID RANGE”

message with the RESET key.
Turning off and on again the humidity switch (4) or turning off and on again the chamber with the main

power switch (3) will reset the flashing notification “HUMID SYSTEM”.