BINDER KBF LQC 240 User Manual

Page 54

Advertising
background image

KBF LQC (E5.3) 07/2014

page 54/114


Program entry as set-point step (example of a temperature program)

W/°C

t/min.

01 02 03 04 05 06 07 08

0

20

40

60

80

100

30

120

180

380

Tolerance

minimum

Tolerance

maximum

Program table corresponding to the diagram above:

Program

section

Set-point

temp.

Fan Section time Operation

lines

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

40

100 % 00:30:00

Off

1

0

-1999

+9999

02

40

100 % 00:00:01

Off

1

0

-1999

+9999

03

60

100 % 01:30:00

Off

1

0

-5

+5

04

60

100 % 00:00:01

Off

1

0

-1999

+9999

05

80

100 % 01:00:00

Off

1

0

-2

+2

06

80

100 % 00:00:01

Off

1

0

-1999

+9999

07

20

100 % 03:20:00

Off

1

0

-1999

+9999

08

20

100 % 00:00:01

Off

1

0

-1999

+9999

Enter now the values of the above program table into the program table of one of the 25 program places

of the MB1 program controller:

For rapid transition phases, do NOT program any tolerance limits in order to permit maximum

heating, refrigerating, and humidification speed.