3 freshwater supply, 1 automatic fresh water supply via water pipe, Caution – BINDER KBF LQC 240 User Manual

Page 27

Advertising
background image

KBF LQC (E5.3) 07/2014

page 27/114

4.3 Freshwater supply

Connect the waste water pipe before connecting the unit to a freshwater pipe or filling the

freshwater can (option, chap. 17.9).

You can supply the unit with freshwater via a water pipe or by manually filling a freshwater can (option,

chap. 17.9).

Water intake temperature NOT below +5 °C / 41°F and not exceeding 40 °C / 104°F.

CAUTION

Calcification of the humidifying system.
Damage to the unit.

Operate the unit with deionized (demineralized) water only.

Types of suitable water quality:

Deionized water from a water treatment installation already existing at the customer's site. Conductivi-

ty from 1 µS /cm up to a maximum of 20 µS/cm. (Water, which is in equilibrium with the CO

2

in the air,

and has a conductivity below 1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH.)

Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable sys-

tem). A reusable measuring equipment to assess the water quality is included (chap. 17.10).

BINDER GmbH is NOT responsible for the water quality at the customer.
Any problems and malfunctions that might arise following use of water of deviating quality is

excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.

4.3.1 Automatic fresh water supply via water pipe

An enclosure inside the unit contains the connection kit for freshwater and wastewater. Install the fresh-

water connection using either the enclosed water hose or another pressure-resistant one. For this re-

move the cover of the freshwater connection “IN” (18) on the rear of the unit. Protect both ends of the

hose with two of the four delivered hose clamps.
Before turning on the unit, check the connection for leaks. Water supply then automatically occurs via the

freshwater connection “IN” (18).

As the appliance only lets in water when required, there is no continuous water flow.

Supply pressure 1 to 10 bar when connecting to a water pipe

Water type: deionized (demineralized) water

Water intake temperature NOT below +5 °C / 41°F and not exceeding 40 °C / 104°F.

The water intake should be provided with a shut-off slide or water-tap.

For the water supply, fix the delivered adapter with hose olive on the thread at the rear of

the chamber.

Protect the water supply at one side with the delivered hose clamp.