BINDER KBF LQC 240 User Manual

Page 99

Advertising
background image

KBF LQC (E5.3) 07/2014

page 99/114


Fault description

Possible cause

Required measures

Humidity
Humidity fluctuation:

Control accuracy of ± 3 % r.F. is

not reached.

Door gasket defective.

Replace door gasket.

Door opened very frequently.

Open doors less frequently.

Humidity fluctuation, together

with temperature fluctuation > 1

°C with a set-point ca. 3 °C above

ambient temperature.

Place of installation too hot.

Select cooler place of installation

or contact BINDER service.

No or too low dehumidification.

Capillary tube blocked

Contact BINDER service.

Not enough refrigerant.

Icing at the evaporator plates.

Set-point for a long time below

ambient temperature.

Defrost the unit (chap. 14).

Condensation at the walls of the

inner chamber.

Combination of tempera-

ture/humidity values not in the

optimum range (see temperature

humidity diagram, Figure 22)

Select combination of tempera-

ture/humidity values in the opti-

mum range (chap. 13).

Set-point for a long time below

ambient temperature, icing in the

preheating chamber.

Defrost the unit (chap. 14)

Bad humidity and temperature

variation

Fan speed has been reduced.

Set fan speed to 100%.

Notification or alarm message

“HUMID SYSTEM” on the con-

troller.

Notification or error of the humidi-

ty system.

See chap. 12.2.

Controller
No unit function

(dark display).

Display mode “Standby” active.

Press any controller key.

Main power switch turned off.

Turn on the main power switch.

No entries to controller keypad

possible. Notification “KEY

LOCK” is displayed

Keyboard locking (option) acti-

vated.

Unlock keyboard locking (chap.

17.7).

No access to menu “User set-

tings”“.

User code incorrect.

Contact BINDER service.

Wrong temperature alarms, dis-

turbance of temperature accuracy Temperature unit changed to °F.

Set temperature unit to °C (chap.

6.4).

Chart recorder function: meas-

ured-value memory cleared, in-

formation lost.

New setting of storage rate.

Change the storage rate ONLY if

the previously registered data are

no longer required (chap. 7).

Controller does not attain set-

points entered in Manual Mode.

Button EXIT or AUTOMATIC has

been hit: Unit is in Idle Mode.

Change to Manual Mode (chap.

8).

Controller does not attain pro-

gram set-points.

Button EXIT or AUTOMATIC has

been hit: Unit is in Idle Mode.

Start the program again (chap.

10.9).

Program duration longer than

programmed.

Tolerances have been pro-

grammed.

For rapid transition phases, do

NOT program tolerance limits in

order to permit maximum heating,

refrigerating, or humidification

speed.

Program stops one section too

early.

Program line is incomplete.

When programming, define the

end value of the desired cycle by

adding an additional section with

a section time of at least one

second.