BINDER KBF LQC 240 User Manual

Page 53

Advertising
background image

KBF LQC (E5.3) 07/2014

page 53/114

10.5 Program entry as set-point ramp or as set-point step

In order to avoid incorrect programming, we recommend plotting the temperature profile and the humidity

profile (chart templates in chap. 10.11 and 10.12) and entering the values into a table (templates in chap

10.13 and 10.14).
The controller provides 8 operation lines (non-functional with standard unit) that can be activated or de-

activated for each program section.

Program entry as set-point ramp (example of a temperature program)

W/°C

t/min.

01

02

03

04

0

20

40

60

80

100

30

120

180

380

05

Tolerance

maximum

Tolerance

maximum

Program table corresponding to the diagram above:

Program

section

Set-point

temp.

Fan Section time Operation

lines

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

40

100 % 00:30:00

Off

1

0

-1999

+9999

02

60

100 % 01:30:00

Off

1

0

-5

+5

03

90

100 % 01:00:00

Off

1

0

-2

+2

04

90

100 % 03:20:00

Off

1

0

-1999

+9999

05

20

100 % 00:00:01

Off

1

0

-1999

+9999

Enter now the values of the above program table into the program table of one of the 25 program places

of the MB1 program controller: