6.2 Function overview of the MB1 program controller displays

The main operation level contains the following differ-

ent displays:

Normal display (Idle Mode or Manual Mode or

Program Mode)

Event List

Chart recorder function

Contact page

Button

VIEW ->

permits toggling between the displays.

The

Normal display enables comparison of the cur-

rent temperature and humidity (W) to the set-point

values (X) or shows the fan working rate.

CONTACT PAGE

BINDER Service contact display.

NORMAL DISPLAY Idle Mode

TEMP

20.0

22.1

°C

HUMID

30.0

21.7

%rH

VIS

0.0

0.0

UVA

0.0

0.0

LIGHT INTEGR. OFF

HAND

08:43:55 15.12.13


or

W

X

08:43:55 15.12.13

CONFIG

VIEW->

PGM

FAN SPEED

50

%

HAND

X - TEMP

26.8

°C

X - HUMID

58.6

%rH

No heating or refrigeration, no humidification or dehumidification. The actual values (X) approximate am-

bient temperature and humidity. Fan working at 50% rate. When the fluorescent tubes are turned on man-

ually, the actual values of light intensity are displayed, light integration is possible.

NORMAL DISPLAY Manual Mode

TEMP

20.0

22.1

°C

HUMID

30.0

21.7

%rH

VIS

0.0

0.0

UVA

0.0

0.0

LIGHT INTEGR. OFF HAND

08:43:55 15.12.13

Temperature and humidity values are maintained

according to the previous entered set-points (W).

The actual light intensity values are displayed. With

integration operating, the samples are automatically

exposed according to the entered target dose values

of UVA and visible spectral range.

NORMAL DISPLAY Program Mode

08:43:55 15.13.13

Program 1 / SEC 3 00 :10 :32

TEMP

20.0

22.1

°C

HUMID

30.0

21.7

%rH

VIS

0.0

0.0

UVA

0.0

0.0

LIGHT INTEGR. OFF AUTO

HAND

A temperature and humidity program entered be-

fore via a program table is run. When the fluores-

cent tubes are turned on manually, the actual val-

ues of light intensity are displayed, light integration

is possible.

08:43:55 15.12.13

Service Hotline