BINDER KBF LQC 240 User Manual

Page 63

Advertising
background image

KBF LQC (E5.3) 07/2014

page 63/114

11.2.2 Checking and setting safety controller set-point type and safety controller set-

point

Grundeinstellung

User-settings

Instrument data

Contrast

Displ. Power down

Contin. operation

User Level

Safety control. Act + 39,8 °C

Safety control. Set. + 42.0 °C

29

User level code no.

1 (factory setting)

User Level

Date and time

Summer time

Language

Temperature unit

Buzzer

Safety controller

User-code No.

English

°C

1

Active

Safety Controller

Setpoint type

Setpoint

+ 42.0

Limit

Sollwert Art

Grenzwer

Offset

Setpoint

+0042.0

Sollwert Art

Grenzwer

Offset

Setpoint type

Limit

Offset

In the menu “User Level” select the submenu “Safety controller”.

Select the safety controller set-point type “Limit” or “Offset” in the field “Setpoint type”

Enter the value for “Limit” or “Offset” in the field “Setpoint”.


For temperature disturbances see alarm indications, chap. 12.

4 x

Advertising