Troubleshooting – BINDER KBF LQC 240 User Manual

Page 98

Advertising
background image

KBF LQC (E5.3) 07/2014

page 98/114

20. Troubleshooting

Fault description

Possible cause

Required measures

Heating

Chamber heating permanently,

set-point not held.

Controller defective.

Contact BINDER service.

Semiconductor relay defective.
Controller not well adjusted, or

adjustment interval exceeded.

Calibrate and adjust controller.

Chamber doesn’t heat up.

Pt 100 sensor defective.

Contact BINDER service.

Heating element defective.
Semiconductor relay defective

Chamber doesn’t heat up when

turning on the chamber.

Safety controller responds.

Limit temperature reached. Safe-

ty controller (chap. 11.2) set too

low.

Let the chamber cool down and

hit RESET button of MB1 control-

ler. If appropriate, select suitable

limit value.

Safety controller (chap. 11.2)

defective.

Contact BINDER service.

Unit without function.

No power supply.

Check connection to power sup-

ply.

Wrong voltage.

Check power supply for voltage

of 115V or 230V.

Unit fuse has responded.

Check unit fuse and replace it if

appropriate. If it responds again,

contact BINDER service.

Controller defective.

Contact BINDER service.

Nominal temperature exceeded

by 10° due to unit failure. Over

temperature protective device

(class 1) responds.

Mechanical safety device class

3.1 responds

(with option safety device class

3.3).

Limit temperature reached.

Check setting of temperature set-

point and of safety device class

3.1. If appropriate, select suitable

limit value.

Too much external heat load.

Reduce heat load.

Controller defective.

Contact BINDER service.

Safety device defective.
Semi-conductor relay defective

Mechanical safety device class

3.2 responds

(with option safety device class

3.3).

Limit temperature reached.

Check setting of temperature set-

point and of safety device class

3.2. If appropriate, select suitable

limit value.

Controller or safety device defec-

tive.

Contact BINDER service.

Refrigerating performance

No or too low refrigerating per-

formance.

Ambient temperature > 25 °C /

77°F (chap.3.4).

Select cooler place of installation.

Combination of tempera-

ture/humidity values not in the

optimum range (see temperature

humidity diagram, Figure 22).

Select combination of tempera-

ture/humidity values in the opti-

mum range (chap. 13).

Compressor not turned on.

Contact BINDER service.

Electro-valves defective.
No or not enough refrigerant.
Too much external heat load.

Reduce heat load.