KBF LQC (E5.3) 07/2014

page 50/114







To enter values, select the corresponding line

via the arrow keys.
Hit the ENTER button. The

program editor

appears.

Enter the individual values of the selected program section.

Program editor Abs.Nr. 6

****.*

Setpoint 1

+100.0

FAN

Operating contacts

Time

00:45:00

Repeat Section

5

Repeat Number

10

Tol.-band min.

-1999.0

Tol.-band max.

+9999.0

Parameter set

1

Setpoint 1

Temperature with start of the program section

FAN

Fan speed in %

Operating contacts Operation lines (no function)
Time

Duration of the program section

Repeat Section

No. of start section in case of repeat cycles

Repeat Number

No. of repeats in case of repeat cycles

Tol.-band min.

Temperature limits (maximum / minimum temperature)

(In case of exceeding: temporary program stop)

Tol.-band max.
Parameter set

Pre-selected value (Do NOT change!)

Select the parameters via the arrow keys and confirm by ENTER.
Then enter the values via the arrow keys, and confirm the entry by ENTER.

For a negative set-point entry, enter the numerical value first, and only then the negative sign

(-).

With set-point type “

Limit”, adapt the safety controller (chap. 11.2) or the temperature safety

device class 3.3 (option, chap. 11.3) to the highest temperature set-point value of the program

actually used. Check the safety device for each temperature program and adapt it if neces-

sary. Set the safety controller set-point or the set-point of temperature safety device class 3.3

(option) by approx. 2 °C to 5 °C above the controller temperature set-point.
Recommended setting: Set-point type “

Offset” with set-point 2 °C.

PGM

08:43:55 15.02.02

PGM

08:43:55 15.12.13