4 kbf lqc technical data – BINDER KBF LQC 240 User Manual

Page 102

Advertising
background image

KBF LQC (E5.3) 07/2014

page 102/114

21.4 KBF LQC technical data

Unit size

240

720

Exterior dimensions
Width

mm / inch

925 / 36.42 1250 / 49.21

Height (incl. castors)

mm / inch

1460 / 57.48 1925 / 75.79

Depth

mm / inch

880 / 34.65

887 / 34.92

Depth including door handle, I-triangle, connection

mm / inch

850 / 33.46

939 / 36.97

Wall clearance rear (minimum) (spacers)

mm / inch

100 / 3.94

100 / 3.94

Wall clearance side (minimum)

mm / inch

100 / 3.94

200 / 7.87

Steam space volume

l / cu.ft.

348 / 12.3

918 / 32.4

Quantity of doors

1

2

Quantity of inner glass doors

1

2

Interior dimensions
Width

mm / inch

650 / 25.60

973 / 38.31

Height

mm / inch

785 / 30.91 1250 / 49.21

Depth

mm / inch

485 / 19.09

576 / 22.68

Interior volume

l / cu.ft.

247 / 8.7

700 / 24.7

Quantity of racks (standard / max.)

2/7

3/12

Quantity of light cassettes

2

3

Load per rack

kg / lbs.

30 / 66

45 / 99

Permitted total load

kg / lbs.

100 / 220

150 / 331

Weight (empty)

kg / lbs.

215 / 474

376 / 829

Temperature data (without humidity)
Temperature range without light cassettes 1)

°C / °F

0 to +70 /

32 to 158

0 to +70 /

32 to 158

Temperature range with light cassettes, with illumination 1)

°C / °F

+10 to +60 /

50 to 140

+10 to +60 /

50 to 140

Max. heat compensation up to 40 °C / 104°F, with illumina-

tion

W

400

1000

Climatic data (with humidity)
Temperature range without light cassettes 1)

°C / °F

+10 to +70 /

50 to 158

+10 to +70 /

50 to 158

Temperature range with light cassettes, with illumination 1)

°C / °F

+10 to +60 /

50 to 140

+20 to +60 /

68 to 140

Temperature fluctuation

with illumination

at 25 °C / 77°F and 60% r.H.

± K

0.2

0.2

at 40 °C / 104°F and 75% r.H.

± K

0.2

0.2

Temperature uniformity

(variation) with illumination

at 25 °C / 77°F and 60% r.H.

± K

0.6

1.2

at 40 °C / 104°F and 75% r.H.

± K

0.6

1.2

Humidity range without light cassettes

% r.H.

10 to 80

10 to 80

Humidity range with light cassettes, with illumination

% r.H.

10 to 75

10 to 75

Humidity fluctuation with

illumination

at 25 °C / 77°F and 60% r.H.

± % r.H.

1.5

≤ 2.0

at 40 °C / 104°F and 75% r.H.

± % r.H.

2

≤ 2.0

Recovery time after doors

were open for 30 sec. 2)

at 25 °C / 77°F and 60% r.H.

minutes

3

1

at 40 °C / 104°F and 75% r.H.

minutes

4

5

Illumination data per light cassette

ICH compliant illumination for photo stability testing 3)

Lux

8000

9000

UVA W/m

2

1.2

1.6

Advertising
This manual is related to the following products: