3 performance after power failure in manual mode – BINDER KBF LQC 240 User Manual

Page 45

Advertising
background image

KBF LQC (E5.3) 07/2014

page 45/114

For a negative set-point entry, enter the numerical value first and only then the negative sign

(-).

When operating without humidity (humidity switch (4) OFF) set the humidity set-point in Man-

ual Mode to 0 % r.H. in order to avoid alarms (in case of the humidity deviating by more than

+/- 5% from the set-point).

8.2 Entering the set point values when operating without illumination

The KBF LQC has been adjusted for operation with illumination. Since the illumination creates a heat

input in the chamber, this must be considered when entering a setpoint without illumination. For this rea-

son, adapting the setting of the temperature set-point according to the following table is required.

Target temperature value

Set-point entry without illumination

KBF LQC 240

KBF LQC 720

10 °C

10 °C

11 °C

20 °C

21 °C

21 °C

30 °C

31 °C

32 °C

40 °C

41 °C

42 °C

50 °C

51 °C

53 °C

60 °C

62 °C

63 °C

70 °C

72 °C

74 °C

Adapting the setting of the humidity set-point according to the following table is also required.

Target humidity value

Set-point entry without illumination

KBF LQC 240

KBF LQC 720

10 % r.H.

10 % r.H.

9 % r.H.

20 % r.H.

19 % r.H.

17 % r.H.

30 % r.H.

26 % r.H.

25 % r.H.

40 % r.H.

36 % r.H.

34 % r.H.

50 % r.H.

45 % r.H.

43 % r.H.

60 % r.H.

55 % r.H.

52 % r.H.

70 % r.H.

65 % r.H.

61 % r.H.

80 % r.H.

75 % r.H.

71 % r.H.


8.3 Performance after power failure in Manual Mode

In Manual Mode (HAND), all functions reinstate exactly to the same situation the chamber had before

power failure. The set-points are immediately resumed. The power failure will appear in the event list. No

error message indicating that a power failure has taken place is displayed.