2 entry of temperature values and fan speed – BINDER KBF LQC 240 User Manual

Page 49

Advertising
background image

KBF LQC (E5.3) 07/2014

page 49/114

10.2 Entry of temperature values and fan speed

08:43:57 15.12.13

Prog. Select.

Prog 1 PROG 01

Prog 2 PROG 02

Prog 3 PROG 03

Prog 4 PROG 04

Prog 5 PROG 05

Prog 6 PROG 06

Prog 7 PROG 07

Prog 8 PROG 08

Prog 9 PROG 09

Prog 10 PROG 10

Prog 11 PROG 11

Prog 1 PROG 01

TP- Program 2

TP- Program 3

TP- Program 1

Select the first subroutine “

TP-Program 1” and confirm with ENTER.

An initially empty

program table appears. You can now enter the temperature program.

Temperature at the

beginning of the

program section


Fan speed in %
Factory setting:

50% in Idle Mode

100% in Manual Mode and

Program Mode

Program No.

Subprogram TP-Program No. 1

Total number of program

sections

Parameter set (preselected)

Tolerance band limits tem-

perature (maximum and mini-

mum temperature)

Number of repeats in case of

repeat cycles

Operation lines

(no function)

Duration of program

section

Number of start section in case

of repeat cycles

You can enter

Program sections into this program table.


Hit the PGM button. An inquiry display appears offering to enter or delete individual program sections.

ZP-Abschnitt Abs. Nr. 5

insert

delete

In this view, new program lines can be entered or deleted:

new

New lines are added below in the table

insert

New lines are added above a previously selected line

delete

Individual lines that have been selected previously are deleted

Create as many lines, i.e. program sections, as desired. In a next step, values can be entered into these

lines. It is as well possible to add supplementary lines later on or to delete individual lines at any moment.

08:43:55 15.12.13

Advertising