2 decontamination, Ole_link4, Ole_link6 – BINDER KBF LQC 240 User Manual

Page 93: Caution

Advertising
background image

KBF LQC (E5.3) 07/2014

page 93/114

18.4.2 Decontamination

The operator must ensure that proper decontamination is performed in case a contamination of the

chamber by hazardous substances has occurred.
Disconnect the chamber from the power supply prior to chemical decontamination. Pull the power plug.
Do not use decontamination agents that may cause a hazard due to reaction with components of the

device or the charging material. If there is doubt regarding the suitability of cleaning products, please

contact BINDER service.
You can use the following disinfectants:

Inner chamber

Standard commercial surface disinfectants free from acid or halides.
Alcohol-based solutions.
We recommend using the disinfectant spray Art. No. 1002-0022.

For chemical disinfection, we recommend using the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from

liability by BINDER GmbH.

With every decontamination method, always use adequate personal safety controls.

In case of contamination of the interior by biologically or chemically hazardous material, there are two

possible procedures depending on the type of contamination and charging material:
1. Spray the inner chamber with an appropriate disinfectant.

Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during

the decontamination process.

2. If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning,

or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.

In case of eye contact, the disinfectant spray may cause eye damage due to chemical

burns. Follow the operating instructions and safety hints labeled on the bottle of the disin-

fectant spray.

Recommended precautions: To protect the eyes use sealed protective goggles.

CAUTION

Eye contact.
Eye damage due to chemical burns.

Do NOT empty into drains.

Wear protective goggles.

After using the disinfectant spray, allow the unit to dry thoroughly, and aerate it sufficiently.