KBF LQC (E5.3) 07/2014

page 33/114

5. Start up

After connecting the supply lines (chap. 4), turn on the unit by the main power switch (3).

Open the water-tap for freshwater supply. Alternatively, fill the freshwater can (option, chap. 17.9).

Turn on the humidifying and dehumidifying system with switch (4) (humidity switch ON/OFF).

After turning on the chamber for the first time, or after an interruption of the power supply, the relative

humidity will increase only after a delay of about 20 minutes. During this period, the relative humidity can

drop considerably.
Warming chambers may release odors in the first few days after commissioning. This is not a quality

defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for

one day and in a well-ventilated location.

WARNING: If customer should use a BINDER chamber running in non-supervised continu-

ous operation, we strongly recommend in case of inclusion of irrecoverable specimen or

samples to split such specimen or samples and store them in at least two chambers, if this is

feasible.


5.1 Function overview of the MB1 display program controller

Pilot lamp: Ready for operation

EXIT button (to exit a menu point)

AUTOMATIC button (to start a previously entered program)

ENTER button (to confirm a selection)

Navigation buttons (functions are assigned by the menu)

Rocker switch (2)

Figure 16: MB1 program controller


The MB1 2-channel program controller controls following values inside the constant climate chamber:
Channel 1: Temperature in °C (range without humidity / without light cassettes: 0 °C / 32°F to 70 °C /

158°F)

Channel 2: Relative humidity in % r.H. (range 10 % r.H. to 80 % r.H.)
You can enter the desired set point values in Manual Mode or Program Mode (chap. 5.2) in the display

controller. For the control range of temperature and relative humidity, see chap. 13).
In addition to displaying the actual values of UVA and the visible spectral range, the function “Light Quan-

tum Control” permits cumulative measurement of the light doses. In Manual Mode you can enter target

dose values of UVA and the visible spectral range. When they are reached, the UVA, and cool white fluo-

rescent tubes automatically turn off independently from each other.