4 menu settings in the “user level” menu – BINDER KBF LQC 240 User Manual

Page 39

Advertising
background image

KBF LQC (E5.3) 07/2014

page 39/114

Safety control.Set

The setting of the tolerance limit of the safety controller (chap. 11.2) is dis-

played. You cannot change it in this view.

Safety control.Act

The measuring result of the safety controller (chap. 11.2) is displayed. The

safety controller compares the value measured by a second independent

temperature sensor to the entered tolerance limit.


6.4 Menu settings in the “User Level” menu

User Level

Date and time

Summer time

Language

Temperature unit

Buzzer
Safety controller

User-code No.

English

°C

1

Active

Date and time

Enter the actual date and time to provide the proper measurement records.

Data is displayed in the chart recorder function (chap. 7) of the controller and

will remain stored in case of a power failure.

Summer time

Time is set one hour in advance during the summer time period.
Setting the summer time switch:

Off: No change to summer time occurs

User timed: Beginning and end of summer time can be set individually

Automatic: The summer time arrangement for central Europe is enabled

(summer time from last Sunday of March until last Sunday of October)

Language

Select the menu language as German, English, or French (chap. 6.1).

Temperature unit

Do NOT change the temperature unit from °C to °F.

Buzzer

Audible alarm buzzer
Inactive: No audible alarm will sound if an alarm event happens (chap. 12).
Active: An audible alarm will sound in case of an alarm event (chap. 12).

Safety controller

Enter a safety controller tolerance limit to prevent temperature from exceeding

this setting. For setting, see chap. 11.2.

User-Code No.

Change the password (“user code”) needed to access the “User Level” menu.

Factory default setting +00001.

Make a note of any change in the user password. There is no ac-

cess to this menu without the correct password.