Defrosting at refrigerating operation, Caution – BINDER KBF LQC 240 User Manual

Page 74

Advertising
background image

KBF LQC (E5.3) 07/2014

page 74/114

14. Defrosting at refrigerating operation

BINDER constant climate chambers are very diffusion-proof. To ensure high temperature precision there

is no automatic cyclic defrosting device. The DCT™ refrigerating system largely avoids icing of the evap-

oration plates. Nevertheless, at very low temperatures the moisture in the air can condense on the evap-

orator plates leading to icing.

Always close the door properly.

Operation with temperature set-points > +5 °C / 41°F at an ambient temperature of 25 °C / 77°F:
The air defrosts the ice cover automatically. Defrosting is continually performed.
Operation with temperature set-points < +5 °C / 41°F:
Icing on the evaporator is possible. Defrost the unit manually.

With temperature set-points < +5 °C / 41°F, regularly defrost the unit manually:

Set the humidity to 0 % r.H. and turn on the humidifying system at humidity switch (4).

Set the temperature to 40 °C / 104°F (Manual Mode).

Let the unit operate for about 30 minutes with closed door.

Too much ice on the evaporator is noticeable by reduced refrigerating performance.

When turning off the unit following prolonged refrigerating operation below +5 °C / 41°F, there is danger

of overflowing due to uncontrolled defrosting of icing on the evaporator.

CAUTION

Uncontrolled defrosting of icing on the evaporator.
Danger of overflowing.
After several days of refrigerating operation below +5 °C / 41°F:

Do NOT directly turn off the unit.

Manually defrost the unit (see description above).

Then, shut down the unit at the main power switch (3) and close the tap of the water

supply.