BINDER KBF LQC 240 User Manual

Page 68

Advertising
background image

KBF LQC (E5.3) 07/2014

page 68/114


(continuation)

Notifying sequence

1

2

3

Group Event

Note

(blue field)

Alarm

(red field)

Zero-voltage

relay alarm

outputs (option)

3

X = representing the actual val-

ues of VIS and UVA

LIGHT kLUX W/m

2

X = representing the integrated

values of VIS and UVA

DOSES MLUXh

Wh/m

2

Dose target value LUX reached

LUX DOSE

REACHED

immediately

UVA AND LUX

DOSE REACHED

when also dose target

value UVA is reached

Dose target value UVA reached

UVA DOSE

REACHED

immediately

UVA AND LUX

DOSE REACHED

when also dose target

value LUX is reached

Switch “LIGHT INTEGRATOR” is

in position “OFF”

LIGHT INTEGR. OFF

4

Maximum temperature of 60°C

with light sensor is reached (only

if at least 1 light sensor is

plugged-in)

LIGHT SENSOR 60

°C!

immediately

5

Temperature deviation of more

than +/- 2 °C of the entered set-

point

TEMP RANGE

immediately

TEMP RANGE

after 16 Min.

after 16 min.

Humidity deviation of more than

+/- 5 % r.H. of the entered set-

point

HUMID RANGE

immediately

HUMID RANGE

after 16 min.

after 16 min.

6

Humidity system turned off with

switch (4)

HUMID OFF

immediately

Door open

DOOR OPEN

immediately

DOOR OPEN

after 5 min.

7

With option locking of the keyboard (BINDER Individual, chap. 17.7):

Locked keyboard

KEY LOCK

immediately

Power failure

immediately

The indicated intervals refer to the time after occurrence of the error or notified condition.
Notifying messages from one group (e.g. Group 2) mask the notifying messages of the subsequent

Group (e.g. Group 3). The same is true for the alarm messages of different groups.
Notifying, and alarm messages of Group 3 are displayed in intervals, so messages of subsequent groups

can appear for a moment alternately.

Advertising
This manual is related to the following products: