3 mounting with wastewater recycling, Caution – BINDER KBF LQC 240 User Manual

Page 88

Advertising
background image

KBF LQC (E5.3) 07/2014

page 88/114

17.9.3 Mounting with wastewater recycling

With a clean chamber interior you can reuse the wastewater from the unit. Connect the wastewater

connection of the chamber (17) with the freshwater hose connection (24) of the freshwater can. The

wastewater can is not used in this case.

CAUTION

Soiling of the vapor humidification system.
Damage to the unit.

Reuse wastewater ONLY with a clean chamber interior.

In case of soiling / contamination of the interior, conduct the wastewater to the waste

water connection or use the wastewater can.


(1) Fixing of the freshwater can (if required)

Hang the can with its holding device on its 4 carriers. You can install it either at the left or the right

side.

(2) Cable connections of the freshwater can

Connect the plug of the cable to the socket (22) at the rear of the unit as described in chap. 17.9.1.

(3) Hose connections

Plug the wastewater hose to the hose connection (24) of the freshwater can and secure it with a

hose clamp. You can use a part of the regularly delivered water hose.
Plug the free hose edge to the wastewater connection “OUT” (17) at the rear of the unit and secure

it with a hose clamp.

Freshwater hose connection (24)

Figure 37: Freshwater can (option)

Bringing a source of humidity into the inner chamber may increase wastewater production.

Regularly check the filling level of the freshwater can.