11.3 Temperature safety device class 3.3 (DIN 12880) (option)

With the option over/under temperature protective device (temperature safety device class 3.3 acc. to

DIN 12880) the unit is equipped with two additional safety devices (class 3.1 and class 3.2). The combi-

nation of the safety devices is regarded as a safety device class 3.3.
The temperature safety device, class 3.3, serves to protect the constant climate chamber, its environment

and the contents against forbidden temperature excesses. Please observe the guideline BGI/GUV-I 850-

0 on safe working in laboratories (formerly BGR/GUV-R 120 or ZH 1/119 laboratory guidelines issued by

the employers’ liability insurance association) (for Germany).
With

safety device class 3.1 a maximum value for the temperature is set that the unit will not exceed

due to the regulatory function of the safety device class 3.1. This protection against excessively high

temperatures can, for example, serve to protect the constant climate chamber, its environment and the

material under treatment from excess temperatures.
With

safety device class 3.2 a minimum value for the temperature is set that the unit will not fall below

due to the regulatory function of the safety device class 3.2. This protection against excessively low tem-

peratures can, for example, serve to protect sensitive loads from undercooling.
Both safety devices are functionally and electrically independent of the temperature control system. If an

error occurs, they perform regulatory function.
Safety devices class 3.1 (14) and class 3.2 (15) are located in the left lateral control panel.

With option temperature safety device class 3.3, the safety controller (chap. 11.2) must be set

to limit 100 °C / 212°F.


(15)

(14)




Excess

temperature

Set-point


Insufficient

temperature




W

X

CONFIG

HAND

VIEW->

EXIT

W

08:43:55 20.09.02

TEMP

°C

40.0

36.8

HUMID

%rH

65.0

58.6

(1)

Figure 19: Temperature safety device class 3.3