2 temperature safety device class 3.2 – BINDER KBF LQC 240 User Manual

Page 66

Advertising
background image

KBF LQC (E5.3) 07/2014

page 66/114

11.3.2 Temperature safety device class 3.2

The safety device class 3.2 is equivalently set to a minimum temperature the

unit will not fall below. This protection against forbidden low temperatures

can for example serve to protect sensitive cultures from cooling down too

much.
If the control knob (15) is turned to its minimum (position 1), the safety de-

vice class 3.2 is without effect. If it is set to a temperature somewhat lower

than that selected by means of the controller, it functions as a protective

device for the material under treatment.

If the temperature safety device class 3.2 has assumed regulation, identifiable by the red alarm lamp

(15a) lighting up, the message “TEMP ALARM” on the controller display, and the buzzer, please proceed

as follows:

Reset the buzzer with RESET key on the controller.

Disconnect the unit from the power supply.

Have examined and rectified the cause of the fault by an expert.

Start up the unit again as described in chap. 5.

Setting:
To check the response temperature of the safety device class 3.2, put the unit into operation and set the

desired set point at the temperature controller.
The sections of the scale from 1 to 10 correspond to the temperature range from -40 °C / -40°F to +160

°C / 320°F and serves as a setting aid.

Turn the control knob (15) of the safety device by means of a coin to position

1 (thermostat without effect).

When the set point is reached, reset the safety device to its trip point (turn it

clockwise).

The trip point is identifiable by the red alarm lamp (15a), the message “TEMP

ALARM” on the controller display, and the buzzer. Reset the buzzer with

RESET key on the controller.

The optimum setting for the safety device is obtained by turning the control

knob anti-clockwise by approximately two scale divisions, which leads to ex-

tinguish the red alarm lamp (15a).

Figure 21: Setting

safety device class 3.2

Check the setting regularly and adjust it following changes of set-point or charge.

Function check:
Check the temperature safety device class 3.2 at appropriate intervals for its functionality. It is recom-

mended that the authorized operating personnel should perform such a check, e.g., before starting a

longer work procedure.

(15a)

(15)

Advertising
This manual is related to the following products: