Danger – BINDER KBF LQC 240 User Manual

Page 82

Advertising
background image

KBF LQC (E5.3) 07/2014

page 82/114

17.4 Analog outputs for temperature and humidity (option)

With this option the chamber is equipped with analog outputs 4-20 mA for temperature and humidity.

These outputs allow transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (6) in the right lateral control panel as follows:

ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –
PIN 2: Temperature +
PIN 3: Humidity –
PIN 4: Humidity +
Humidity range: 0 % r.H. to 100 % r.H.
Temperature range: -10 °C / 14°F to +100 °C / 212°F
A suitable DIN plug is enclosed.

Figure 29: Pin allocation of DIN socket (6) for option analog outputs

17.5 Zero-voltage relay alarm outputs for temperature and humidity for alarm

transmission (option)

Equipment of the chamber with zero-voltage relay outputs for temperature and hu-

midity (option) permits the transmission of alarms to a central monitoring system.

Connection is established via a DIN socket.

Figure 30: Pin configuration of the DIN socket

Temperature contact

Humidity contact

Pin 1: Pin
Pin 2: Make

Pin 3: Pin
Pin 4: Make

In case of a temperature alarm, pins 1 and 2 are open; with humidity alarm, pins 3 and 4 are open. This

happens simultaneously to the alarm message displayed at the controller display.
In case of power failure, both contacts are open.
Maximum loading capacity of the switching contacts: 24V AC/DC - 2,5A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.

Do NOT connect any devices with a higher loading capacity.

In case of temperature and humidity alarm, the alarm message on the controller display remains dis-

played during transmission of an alarm via the zero-voltage relay outputs.
As soon as the cause of the alarm is rectified, you can reset the alarm transmission via the zero-voltage

relay outputs together with the alarm message at the controller display hitting the RESET key.

3

4

1
2