Notification and alarm functions – BINDER KBF LQC 240 User Manual

Page 67

Advertising
background image

KBF LQC (E5.3) 07/2014

page 67/114

12. Notification and alarm functions

12.1 Notification and alarm system overview (auto diagnosis system)

The unit provides notification and alarm functions, which indicate messages in up to three steps:
1. Visual indication of notifying or error messages are blue notes on the display of the MB1 controller.
2. Visual indication of alarm messages are red notes with an alarm bell symbol. After a delay time, some

notes change their color from blue to red.

In addition, there is an audible alert, if you did not deactivate the buzzer in the “User level” menu

(chap. 6.4).

3. Zero-voltage relay outputs (option, chap. 17.5) allow transmitting the alarm e.g., to a central monitor-

ing system.

Notifying sequence

1

2

3

Group Event

Note

(blue field)

Alarm

(red field)

Zero-voltage

relay alarm

outputs (option)

1

Exceeded temperature limit of

the safety controller

TEMPERATURE

LIMIT

immediately

Error of the humidity system.

Press RESET key for further

information.

HUMID SYSTEM

immediately

HUMID SYSTEM

after 1 min.

Notification or error of the humidi-

ty system.
For detailed information see

chap. 12.2

HUMID SYSTEM

(notification flashes)

2

With option temperature safety device class 3.3 (see chap. 11.3):
Exceeding the maximum or min-

imum temperature

TEMP ALARM

immediately

Advertising
This manual is related to the following products: