To set the configuration – RIGOL DM3068 User Manual

Page 121

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

User’s Guide for DM3068

2-79

To Set the Configuration


Press

 System  Cfg to enter the following interface.

1. PwrOn

Select a system configuration to be used at power-on from “Default” and
“Last (configuration at last power-off).

 This setting will be available at the next power-on.
 The multimeter will always use DCV as the default measurement at every

power-on no matter Default or Last is selected.

2. Default

Restore the multimeter to factory defaults (see Table 2-8).

3. Switch

Enable or disable the front power button. The default status is OFF.

ON: enable the front power button. After turning on the power switch at

the rear panel, users need to press the power button at the front panel to
turn on the multimeter.

OFF: disable the front power button. Users can directly start the

multimeter by turning on the power switch at the rear panel. However,
you can still turn off the multimeter by pressing the power button at the
front panel.


Table 2-8 Factory defaults

Parameter

Factory Default

Measurement Parameters
*Short-circuit Resistance

10 Ω

AC Filter

Mid

DC Input Impedance

10 MΩ

Reading Resolution

6 ½

Measurement Function

DCV