Loading errors (dcv), Loading errors (dcv) -2 – RIGOL DM3068 User Manual

Page 142

Advertising
background image

RIGOL

Chapter 5 Measurement Tutorial

User’s Guide for DM3068

5-2

Loading Errors (DCV)


Measurement loading errors occur when the resistance of the device-under-test
(DUT) takes up a certain percentage in the multimeter’s own input resistance. The
diagram below shows this error source.

HI

LO

Rs

Vs

Ri

Ideal

Meter

Vs = ideal DUT voltage
Rs = DUT source resistance
Ri = input resistance of the multimeter (10 MΩ or >10 GΩ)

Error (%) =

i

R

s

R

s

R

100

+

×


To reduce the effect of loading errors and noise interference, you can set the input
resistance of the multimeter to “>10GΩ” for 200 mV, 2 V and 20 V ranges, and
“10MΩ” for the 200 V and 1000 V ranges.













Advertising
This manual is related to the following products: