Dc impedance, Dc impedance -42 – RIGOL DM3068 User Manual

Page 84

Advertising
background image

RIGOL

Chapter 2 Front Panel Operations

User’s Guide for DM3068

2-42

DC Impedance


DC impedance applies to DCV measurement. The default is “10MΩ” and is stored
in non-volatile memory. In the range of 200 mV, 2 V or 20 V, you can choose
“>10GΩ” to reduce the loading error to the measured object caused by the
multimeter (refer to Loading Errors (DCV)”). The current selection is stored in
volatile memory.


In DCV measurement, press

 Res to enter the following interface.


Press 10MΩ or >10GΩ to select the desired impedance.
10MΩ: set the input impedances in all ranges to 10 MΩ.
>10GΩ: set the input impedances in ranges of 200 mV, 2 V and 20 V to >10

GΩ, while in ranges of 200 V and 1000 V, the impedances are still 10 MΩ.