Chapter 4 Troubleshooting

RIGOL

User’s Guide for DM3068

4-1

Chapter 4 Troubleshooting


The commonly encountered failures and their solutions are listed below. When you
encounter those problems, please solve them following the corresponding steps. If
the problem remains still, please contact

RIGOL and provide your device

information (

T/C Info).


1. The screen is still dark (no display) after power on.

1) Check whether the power cord is well connected.
2) Check whether the power switch on the rear panel is switched on.
3) Check whether the fuse is burned out. If the fuse needs to be changed,

please use the specified fuse.

4) Restart the instrument after finishing the above inspections.
5) If the problem persists, please contact

RIGOL.


2. The reading does not change when a current signal is input.

1) Check whether the test lead is correctly inserted into the HI and LO

terminals of current measurement.

2) Check whether the current fuse at the rear panel is burned out.
3) Check whether the DCI or ACI measurement function is enabled.
4) Check whether the DCI measurement function is used to measure AC

current.


3. The reading is abnormal when a DC signal is input.

1) Check whether the test lead is correctly inserted into the HI and LO

terminals of current measurement.

2) Check whether the current fuse at the rear panel is burned out.
3) Check whether the DCI or DCV measurement function is enabled.
4) Check whether the ACI measurement function is used to measure DC

current.


4. The USB storage device cannot be identified.

1) Check whether the USB storage device is in good condition.
2) Make sure the USB storage device you used is a flash storage device. This

instrument does not support hardware storage type.

3) Check the capacity of your USB storage device. It is recommended that