Frequency and period characteristics, Frequency and period characteristics -11 – RIGOL DM3068 User Manual

Page 157

Advertising
background image

Chapter 6 Specifications

RIGOL

User’s Guide for DM3068

6-11

Frequency and Period Characteristics

Accuracy Specifications: ±(% of reading)

[1][2]

Function

Range

Frequency Range

24 Hour

[3]

T

CAL

±1

90 Day
T

CAL

±5

1 Year
T

CAL

±5

Temperature
Coefficient
0
to (T

CAL

-5)

(T

CAL

+5)to 50

Frequency, Period

200 mV to
750 V

3 Hz-5 Hz

0.07

0.07

0.07

0.005

5 Hz-10 Hz

0.04

0.04

0.04

0.005

10 Hz-40 Hz

0.02

0.02

0.02

0.001

40 Hz-300 kHz

0.005

0.006

0.007

0.001

300 kHz-1 MHz

0.005

0.006

0.007

0.001

Additional Low Frequency Errors: (% of reading)

Frequency

Gate Time (Resolution)
1 s (0.1 ppm)

0.1 s (1 ppm)

0.01 s (10 ppm)

0.001 s (100 ppm)

3 Hz-5 Hz

0

0.12

0.12

0.12

5 Hz-10 Hz

0

0.17

0.17

0.17

10 Hz-40 Hz

0

0.20

0.20

0.20

40 Hz-100 Hz

0

0.06

0.21

0.21

100 Hz-300 Hz

0

0.03

0.21

0.21

300 Hz-1 kHz

0

0.01

0.07

0.07

>1 kHz

0

0

0.02

0.02

[1]

Specifications are for 90 minutes warm-up and 1 s gate time.

[2]

For frequency ≤300 kHz, the specification is for AC input voltage of 10% to 110% of range. For frequency >300 kHz, the specification is for AC input
voltage of 20% to 110% of range. The maximum input is limited to 750 V rms or 8 x 10

7

Volts-Hz (whichever is less). 200 mV range is full range input

or input that is larger than the full range. For 20 mV to 200 mV inputs, multiply % of reading error by 10.