Temperature characteristics, Temperature characteristics -14 – RIGOL DM3068 User Manual

Page 160

Advertising
background image

RIGOL

Chapter 6 Specifications

User’s Guide for DM3068

6-14

Temperature Characteristics

Accuracy Specifications

[1]

Function

Probe Type

Type

Optimum Range

1 Year
T

CAL

±5

Temperature Coefficient
0
to (T

CAL

-5)

(T

CAL

+5) to 50

Temperature

RTD

[2]

(R

0

is within 49 Ω

and 2.1 kΩ)

α=0.00385

-200℃ to 660℃

0.16℃

0.01℃

α=0.00389

-200℃ to 660℃

0.17℃

0.01℃

α=0.00391

-200℃ to 660℃

0.14℃

0.01℃

α=0.00392

-200℃ to 660℃

0.15℃

0.01℃

Thermal
Resistance

2.2 kΩ

-40℃ to 150℃

0.08℃

0.002℃

3 kΩ

-40℃ to 150℃

0.08℃

0.002℃

5 kΩ

-40℃ to 150℃

0.08℃

0.002℃

10 kΩ

-40℃ to 150℃

0.08℃

0.002℃

30 kΩ

-40℃ to 150℃

0.08℃

0.002℃

Thermocouple

[3]

B

0℃

to 1820℃

0.76℃

0.14℃

E

-270℃ to 1000℃

0.5℃

0.02℃

J

-210℃ to 1200℃

0.5℃

0.02℃

K

-270℃ to 1372℃

0.5℃

0.03℃

N

-270℃ to 1300℃

0.5℃

0.04℃

R

-270℃ to 1768.1℃

0.5℃

0.09℃

S

-270℃ to 1768.1℃

0.6℃

0.11℃

T

-270℃ to 400℃

0.5℃

0.03℃

[1]

Specifications are for 90 minutes warm-up. Probe error excluded.

[2]

Specification is for 4WR resistance measurement or 2WR resistance measurement using REL operation for sensor.

[3]

Relative to cold junction temperature, accuracy is based on ITS-90. Built-in cold junction temperature refers to the temperature inside the banana jack
and its accuracy is ±2.5 ℃.

Advertising
This manual is related to the following products: