General specifications, General specifications -20 – RIGOL DM3068 User Manual

Page 166

Advertising
background image

RIGOL

Chapter 6 Specifications

User’s Guide for DM3068

6-20

General Specifications

Display

256×64 LCD, dual display, Chinese/English menu and online help.

Power Supply

AC 100 V – 120 V, 45 Hz – 440 Hz
AC 200 V – 240 V, 45 Hz – 66 Hz
Detect the power frequency automatically at power-on, 400 Hz defaults to 50 Hz

Power Consumption

25 VA Max

Working Environment

Full accuracy for 0℃ to 50℃
Full accuracy to 80% R.H. at 40℃ Non-coagulation

Storage Temperature

-40℃ to 70℃

Operation Altitude

Up to 2000m

Safety

IEC 61010-1; EN 61010-1; UL 61010-1; CAN/CSA-C22.2 No. 61010-1
Measurement CAT I 1000V/CAT II 300V
Pollution Degree 2

EMC

EN 61326-1

Weight

About 3.2 kg (without package)

Dimension

(height×width×length): 107.0 mm×231.6 mm×290.5 mm

Remote Interface

GPIB, 10/100Mbit LAN, USB 2.0 Full Speed Device & Host (support USB storage device), RS-232C

Programming Language

SCPI

LXI Compatibility

LXI Class C, Version 1.2

Warm-up Time

90 minutes