Chapter 6 Specifications

User’s Guide for DM3068

6-8

10.00000 A

[6]

3 Hz-5 Hz

1.10 + 0.08

1.10 + 0.10 1.10 + 0.10

0.100 + 0.008

5 Hz-10 Hz

0.35 + 0.08

0.35 + 0.10 0.35 + 0.10

0.035 + 0.008

10 Hz-5 kHz

0.15 + 0.08

0.15 + 0.10 0.15 + 0.10

0.015 + 0.008

Additional Low Frequency Errors (% of reading)

Additional Crest Factor Errors (non-sine wave)

[7]

Frequency

AC Filter

Crest Factor

Error (% of reading)

Slow

Medium

Fast

10 Hz-20 Hz

0

0.74

--

1 - 2

0.05

20 Hz-40 Hz

0

0.22

--

2 - 3

0.2

40 Hz-100 Hz

0

0.06

0.73

3 - 4

0.4

100 Hz- 200 Hz

0

0.01

0.22

4 - 5

0.5

200 Hz-1 kHz

0

0

0.18

>1 kHz

0

0

0

[1]

Specifications are for 90-minute warm-up, slow ac filter and sinewave input.

[2]

10% overrange on all ranges except ACV 750 V and ACI 10 A ranges.

[3]

Relative to calibration standards.

[4]

Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50 kHz
to 100 kHz, add 0.13% of range.

[5]

ACV 750 range limited to 8x10

7

Volt-Hz. When the input exceeds 300 V rms, add 0.7 mV error for each additional volt.

[6]

For continuous current > DC 7A or AC 7A rms, 30 seconds ON and 30 seconds OFF.

[7]

When the frequency is lower than 100 Hz, the specification of slow filter is only for sine wave input.

[8]

Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range, add 0.1% of range additional error. Specifications are typical
values for 200 uA, 2 mA, 2 A and 10 A ranges when frequency is >1 kHz.