Measurement rate, Measurement rate -16 – RIGOL DM3068 User Manual

Page 162

Advertising
background image

RIGOL

Chapter 6 Specifications

User’s Guide for DM3068

6-16

Measurement Rate

Measurement Rate

[1]

Function

Setting

Integration Time

Readings/s 50Hz (60Hz)

DC Voltage
DC Current
2-wire Resistance
4-wire Resistance

0.006 PLC Integration Time

100 (100) us

10000 (10000)

0.02 PLC

400 (333) us

2500 (3000)

0.06 PLC

1.2 (1) ms

833 (1000)

0.2 PLC

4 (3.33) ms

250 (300)

1 PLC

20 (16.7) ms

50 (60)

2 PLC

40 (33.3) ms

25 (30)

10 PLC

200 (167) ms

5 (6)

100 PLC

2 (1.67) s

0.5 (0.6)

AC Voltage
AC Current

[2]

3Hz AC Filter

0.2

20 Hz

1.5

200 Hz

10

200 Hz

50

[3]

Frequency and
Period

[4]

1s Gate Time

1

0.1 s

10

0.01 s

80

0.001 s

500

Capacitance

[5]

25

[1]

Auto trigger, zero trigger delay, auto zero off, auto range off, math function off and external interface off.