RIGOL DM3068 User Manual

Page 150

Advertising
background image

RIGOL

Chapter 6 Specifications

User’s Guide for DM3068

6-4

Measuring Characteristics
DC Voltage
Input Impedance

200 mV, 2 V, 20 V ranges: 10 MΩ or >10 GΩ
(For these ranges, input beyond ±26 V are clamped through 106 kΩ)
200 V and 1000 V ranges: 10 MΩ±1%

Input Protection

1000 V

Input Offset Current

50 pA, at 25℃, typical

CMRR (common mode
rejection ratio)

140 dB for 1 kΩ unbalanced resistance in LO lead, ±500 VDC peak maximum.

Resistance
Measurement Method

4-wire or 2-wire resistance
Current source referenced to LO input

Open-circuit Voltage

Limited to <10 V

Max. Lead Resistance
(4-wire)

10% of range per lead for 200 Ω and 2 kΩ ranges, 1 kΩ per lead on all other ranges

Input Protection

1000 V on all ranges

Offset Compensation

Available on 200 Ω, 2 kΩ and 20 kΩ ranges.

DC Current
Shunt Resistor

100 Ω for 200 uA, 2 mA
1 Ω for 20 mA , 200 mA
0.01 Ω for 2 A, 10 A

Input Protection

Externally accessible 500 mA, 250 V high speed fuse at the rear panel for 200 uA, 2 mA, 20 mA and 200 mA ranges.
Internal 10 A, 250 V high speed fuse for 2 A and 10 A ranges.

Advertising