Energy protocol – ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 134

Advertising
background image

Chapter 4

Supervisor Menu

128

zoll.com

9650-001358-01 Rev. B

Energy Protocol

The following options allow you to set the energy levels of the shocks given.

Adult Shock 1 Energy

The Adult Shock 1 Energy parameter allows you to set the energy level of the first shock given
for adult patients.

The values for the Adult Shock 1 Energy parameter are:

Adult Shock 2 Energy

The Adult Shock 2 Energy parameter allows you to set the energy level of the second shock
given for adult patients.

The values for the Adult Shock 2 Energy parameter are:

Adult Shock 3 Energy

The Adult Shock 3 Energy parameter allows you to set the energy level of the third shock given
for adult patients.

The values for the Adult Shock 3 Energy parameter are:

Pediatric Shock 1 Energy

The Pediatric Shock 1 Energy parameter allows you to set the energy level of the first shock
given for pediatric patients.

The values for the Pediatric Shock 1 Energy parameter are:

Possible Values:

1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J 10 J, 15 J, 20 J, 30 J, 50 J, 70 J 85
J, 100 J, 120J, 150 J, 200 J

Default:

120 J

Possible Values:

1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J 10 J, 15 J, 20 J, 30 J, 50 J, 70 J 85
J, 100 J, 120J, 150 J, 200 J

Default:

150 J

Possible Values:

1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J 10 J, 15 J, 20 J, 30 J, 50 J, 70 J 85
J, 100 J, 120J, 150 J, 200 J

Default:

200 J

Possible Values:

1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J 10 J, 15 J, 20 J, 30 J, 50 J, 70 J 85
J, 100 J, 120J, 150 J, 200 J

Default:

50 J