Alarms (cont’d)

Default Neonate>T2
Alarm (cont’d)

Lower Limit Enable

Off

On, Off

Upper Limit

°F: 100.0

°C: 37.8

°F: 34.0-122.0

°C: 2.0-50.0

Lower Limit

°F: 95.0

°C: 35.0

°F: 32.0-120.0

°C: 0.0-48.0

Default Neonate>

T

Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

°F: 5.0

°C: 2.8

°F: 0.2-90.0

°C: 0.2-50.0

Lower Limit

°F: 0.0

°C: 0.0

°F: 0.0-89.8

°C: 0.0-49.8

Default Neonate>P1
Systolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

100

-28-300

Lower Limit

50

-30-298

Default Neonate>P1
Diastolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

70

-28-300

Lower Limit

30

-30-298

Default Neonate>P1
Mean Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

80

-28-300

Lower Limit

35

-30-298

Default Neonate>P2
Systolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

100

-28-300

Lower Limit

50

-30-298

Default Neonate>P2
Diastolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

70

-28-300

Lower Limit

30

-30-298

Default Neonate>P2
Mean Alarm

Upper Limit Enable

Off

On, Off

Menu

Submenu

Parameter Name

Default Value

Parameter Values