ECG Monitoring Parameters

ECG monitoring parameters allow you to view and modify functions related to using the X
Series unit’s ECG monitoring function. These parameters include:

Filter Response

HR/PR Tone

Resp Monitoring

Resp Lead

Patient Pace Speed

ECG Sweep Speed

Filter Response

The Filter Response parameter allows you to specify to what the ECG bandwidth filter is.

The values for the Filter Response parameter are:

Limited:0.05-40 Hz

Monitor:0.67-40 Hz

EMS Diagnostic:0.05-150 Hz

HR/PR Tone

The HR/PR Tone parameter allows you to enable or disable the heart rate and pulse rate tone on
the X Series unit.

The values for the HR/PR Tone parameter are:

Resp Monitoring

The Resp Monitoring parameter allows you to enable or disable the X Series respiration
monitoring program.

The values for the Resp Monitoring parameter are:

Possible Values:

Limited, Monitor, EMS Diagnostic

Default:

Monitor

Possible Values:

On, Off

Default:

On

Possible Values:

On, Off

Default:

On