ECG Monitoring Parameters
ECG monitoring parameters allow you to view and modify functions related to using the X
Series unit’s ECG monitoring function. These parameters include:
•
Filter Response
•
HR/PR Tone
•
Resp Monitoring
•
Resp Lead
•
Patient Pace Speed
•
ECG Sweep Speed
Filter Response
The Filter Response parameter allows you to specify to what the ECG bandwidth filter is.
The values for the Filter Response parameter are:
Limited:0.05-40 Hz
Monitor:0.67-40 Hz
EMS Diagnostic:0.05-150 Hz
HR/PR Tone
The HR/PR Tone parameter allows you to enable or disable the heart rate and pulse rate tone on
the X Series unit.
The values for the HR/PR Tone parameter are:
Resp Monitoring
The Resp Monitoring parameter allows you to enable or disable the X Series respiration
monitoring program.
The values for the Resp Monitoring parameter are:
Possible Values:
Limited, Monitor, EMS Diagnostic
Default:
Monitor
Possible Values:
On, Off
Default:
On
Possible Values:
On, Off
Default:
On