Supervisor Menu

The following table includes parameter ranges and default values for the Supervisor menu:

Menu

Submenu

Parameter Name

Default Value

Parameter Values

Alarms

General

Default Patient Mode Adult

Adult, Pediatric, Neonate

Volume

Medium

Low, Medium, High

Alarm Suspend Time 4 min

Disable, 2 min, 4 min, 15
min, Always On

Default Adult>LTA
Monitoring

LTA Monitoring

Off

On, Off

Default Adult>HR/PR
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

120

32-300

Lower Limit

50

30-298

Default Adult>NIBP
Systolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

220

32-260

Lower Limit

75

30-258

Default Adult>NIBP
Diastolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

110

22-220

Lower Limit

35

20-218

Default Adult>NIBP
MAP Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

120

25-230

Lower Limit

50

23-228

Default Adult>EtCO2
Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

60

5-150

Lower Limit

8

0-145

Default Adult>FiCO2
Alarm

Upper Limit Enable

Off

On, Off

Upper Limit

8

2-98

Default Adult>RR/BR
Alarm

Upper Limit Enable

Off

On, Off