Supervisor Menu

9650-001358-01 Rev. B

ZOLL Configuration Manual

161

ECG (cont’d)

12 Lead ECG (cont’d) 12-Lead Format

3x4

2x6, 3x4 3x4 Cabrera

Print 10s Ldll

On

On, Off

12-Ld Frequency
Response

Filtered Diagnostic

Diagnostic, Filtered
Diagnostic

12-Ld Analysis
Enabled

On

On, Off

NIBP

Default NIBP Mode

Manual

Manual, Auto

Default Auto Mode
Interval

5 min

1 min, 2 min, 3 min, 5 min,
10 min, 15 min, 30 min, 60
min

Default Cuff Inflation
Target (Adult)

160

120-260; by 20’s

Default Cuff Inflation
Target (Ped.)

120

80-150; by 10’s

Default Cuff Inflation
Target (Neo.)

90

60-130; by 10’s

Printer

Print on Defib/Pacer
Events

On

On, Off

Print on Event

Off

On, Off

Print on Treatment
Snapshot

Off

On, Off

Print Treatment
Summary Snapshots

On

On, Off

Print on Analysis
Results

Off

On, Off

Print on Presenting
Rhythm

Off

On, Off

Display/
Configuration

Localization

Language

English

English, Swedish, Finnish,
Danish, Dutch, Spanish,
German, Italian, French

Decimal Format

.

. (dot), , (comma)

Date Format

MM/DD/YYYY

MM/DD/YYYY, DD/MM/YYYY,
YYYY/MM/DD, MM-DD-YYYY,
DD-MM-YYYY, YYYY-MM-DD

Time Format

24 Hour

12 Hour, 24 Hour

Temp Units

°F

°F, °C

IBP Units

mmHg

mmHg, kPa

CO2 Units

mmHg

mmHg, kPa,%

NIBP Units

mmHg

mmHg, kPa

ECG Lead Labels

AAMI

AAMI, IEC

NIBP Display Format S/D (M)

S/D (M), (M) S/D

Length Units

Inch

Inch, cm

Menu

Submenu

Parameter Name

Default Value

Parameter Values