3.

Select Importing from USB. The message OPERATION WILL OVERWRITE SYSTEM
CONFIGURATIONS AND ALARM SETTINGS. ARE YOU SURE?
displays.

4.

Press YES to complete the import process. The message DO NOT DISCONNECT THE USB
displays. When the import is complete, the message IMPORT IS COMPLETE. YOU MAY
NOW REMOVE THE USB DRIVE. THE DEVICE WILL AUTOMATICALLY RESET AFTER
PRESSING THE “OK” BUTTON
displays.

5.

Press OK.

Note:

Review critical configuration values such as alarm, AED, defibrillator, and pacer
settings to verify that you imported the correct configuration file.

Refer to page 9 for a listing of error messages that may occur while you are importing or
exporting configuration information.