Resp parameter settings, Rr/br alarm – ZOLL X Series Monitor Defibrillator Rev B User Manual
Page 35

Resp Parameter Settings
9650-001358-01 Rev. B
ZOLL X Series Configuration Manual
29
Resp Parameter Settings
The Resp parameter settings allow you to view and modify functions related to using the X
Series unit for respiration monitoring, and include:
•
RR/BR Alarm
•
No Breath Alarm
•
Resp Monitoring
•
Resp Lead
•
CO2/Resp Sweep Speed
RR/BR Alarm
The following options allow you to set the limits of the Respiration Rate alarm for all patient
age groups.
Upper Limit Enable
The Upper Limit Enable parameter allows you to enable or disable the upper limit.
The values for the Upper Limit Enable parameter are:
Lower Limit Enable
The Lower Limit Enable parameter allows you to enable or disable the lower limit.
The values for the Lower Limit Enable parameter are:
Upper Limit
The Upper Limit parameter allows you to set the upper limit.
The values for the Upper Limit are:
Possible Values:
On, Off
Default:
Off
Possible Values:
On, Off
Default:
Off
Possible Values:
5-150
Default:
Adult: 50
Pediatric: 50
Neonate: 80