Resp Parameter Settings

The Resp parameter settings allow you to view and modify functions related to using the X
Series unit for respiration monitoring, and include:

RR/BR Alarm

No Breath Alarm

Resp Monitoring

Resp Lead

CO2/Resp Sweep Speed

RR/BR Alarm

The following options allow you to set the limits of the Respiration Rate alarm for all patient
age groups.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit are:

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

5-150

Default:

Adult: 50

Pediatric: 50

Neonate: 80