Resp lead, Patient pace marker, Ecg sweep speed – ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 19

Advertising
background image

ECG Monitoring Parameters

9650-001358-01 Rev.B

ZOLL X Series Configuration Manual

13

Resp Lead

The Resp Lead parameter allows you to set which electrode the X Series unit monitors as its
respiration lead.

The values for the Resp Lead parameter are:

Patient Pace Marker

The Patient Pace Marker parameter allows you to enable or disable the patient Pace Marker.

The values for the Patient Pace Marker parameter are:

ECG Sweep Speed

The ECG Sweep Speed parameter allows you to set the ECG sweep speed.

The values for the ECG Sweep Speed parameter are:

Possible Values:

Lead I (RA-LA), Lead II (RA-LL)

Default:

Lead II (RA-LL)

Possible Values:

On, Off

Default:

On

Possible Values:

25 mm/s, 50 mm/s

Default:

25 mm/s